DTSSD22英文说明书浩琳达
1.General:
D
SD22three-phasemulti-functionelectronicenergymeterisspecificintegrated,high-techintelligentinstrument.
Themeter,adoptingthestate-of-the-artASIC,non-volatilememory,standardRS485interface,IRcommunication,LCDdisplay,isdesignedandmanufacturedwithSMTtechnicsaccordingtoIECandintermsofChinesesituation.
Thiskindofenergymeterintegratesmanyfunctions,realizingactivebi-directional,reactive4quadrantTOUenergy,splitphase,demandandpowerfactormetering,displayingandremotelytransmittingreal-timevoltage,current,power,loadprofileandetc;meantime,novoltage,phaseloss,currentloss,voltagequalificationratelog,alarmanddisplayfunctionareavailable,whichcaneffectivelypreventstamperingaswellasmeetselectricpowermanagementautomationrequirement.
Uponrequestoftheusers,themetercanbeinstalledwithradiomodulesothatremotemeterreadingchannelisavailable.
1.1Performance
1.1.1Themeter’scircuitdesignandcomponentsadoptingarebasedonalargerangeofenvironmenterror,assuringthemeterbeingabletoworkstablyinalongterm.Basically,theaccuracywillnotbevariousuponfrequency,temperatureandvoltagechanging.Smallsize,lightweight,goodsealingproperty,highreliabilityandstrongresistancetodisturbance
1.1.2ThebackupLithiumbatteryisavailable,itcanguaranteethatthedatawillnotlose,calendar,clockandtimeperiodprocedurecontrolcanworknormallywhenthepowerisoff,andthemeterwillautomaticallyoperatewhenthepowerisrestored.
1.1.3Photo-electricitycouplingpulseoutputportissetonmeterterminalblock,easyforerrortestanddatacollecting.Pulseoutputconstantisconsistentwithmeterconstant.
1.1.4MeterrunninginformationcanbetransmittedthroughIRhandheldcomputer,RS485andphoto-electricityinterface.
1.1.5Theworktemperaturerangeis-40~65℃,andtherelativehumidityshouldnotexceed85%
1.2Workingprinciple:
AftervoltageandcurrentsignalsinPhaseA,B,CareconvertedintocorrespondingdigitalinformationthroughASIC,themeterwillcalculatethevoltage,current,powerandenergyineachphase;CPUreadsthedatathroughSPI,obtainsandrecordsenergy,demandandpowerfactoretc.throughprocedureprocessing.Atthesametime,itcanalsodistinguishtheabnormityofvoltageorcurrentineachphaseandrecordthestatusofcorrespondingvoltagelossandcurrentloss.
1.3SpecificationandMainTechnicalparameter
AlltheindexmeetstandardofIEC62053.
1.3.1Specification
Model
Accuracy
RatedVoltage(V)
RatedCurrent(A)
DSSD22
Active0.2S/0.5S/1.0
Reactive2.0
3×100
3×380
0.3(1.2)1.5(6)3(6)
5(10)5(20)10(40)20(80)20(100)30(100)
DTSD22
Active0.2S/0.5S/1.0
Reactive2.0
3×57.7/100
3×220/380
3×230/400
0.3(1.2)1.5(6)3(6)5(10)5(20)10(40)20(80)20(100)30(100)
Uponrequestoftheuser,ourcompanywilldesignandproducemeterswithallkindsofspecialvoltage,current,functionandcommunicationprotocoltorealizeenergymeteringonspecialcondition.
D
SD22SS1importandexportaccuracyarebothsatisfiedwithaccuracyrequirement,andimportandexportaccuracyoperateatthesametimeperiod.
D
SD22SS2importandexportaccuracyarebothsatisfiedwithaccuracyrequirement,whileexportandimportoperateattheirowntimeperiod.
D
SD22SD1saveexportdataandexportenergyfortheuser’sreference,andimportandexportaccuracyoperateatthesametimeperiod.
D
SD22SD2saveexportdataandexportenergyfortheuser’sreference,andexportandimportaccuracyoperateatthesametimeperiod.
1.3.2MainTechnicalParameters:
Clockerror:
±0.5s/d(temperaturecompensationhardclock)
Powerconsumption:
LCDdisplay,voltagecircuit≤1.2W/5VA,currentcircuit≤0.5VA
Voltagerangeforpowersupply:
70%Un~130%Un
Backuppowersupply:
doublelithiumbattery,3.6v1.2Ah,thedatacanbesavedforover10years
Thelifeofbattery:
≥10years
Accuracy:
activeclass0.2S/0.5S/1.0,reactiveclass2.0
Startingcurrent:
≤0.1%Ib(0.2S,0.5S),≤0.2%Ib(Class1.0)
Runningwithnoload:
logicalcircuitagainstrunningwithnoloadisavailable
Tariff&Period:
Themeterhas32energytariffregistersand24demandtariffregisters.10setstimezonescanbeseteveryyear.4tariffs,12timeperiods,12holidayandweekendcanbeseteveryday.Alsotheusercansetdifferenttimezones,timeperiod,andnumberoftariffuponrequirement.
1.4Adjustmentandcalibration
1.4.1Adjustment
Beforeoutgoingthefactory,themeterisadjustedandsealed,soadjustmentisnotrequiredafteroutgoing.Ifadjustmentneeded,itcanbedonethroughsoftwareprovidedbyourcompany.
1.4.2Calibration
Thiskindofwatt-hourmeterhasspecifictestingapertureforactiveandreactive.Whenyoutestit,inordertoensurethetestingaccuracy,youshouldaddthepulsewhenthecurrentincreasesatthedifferenttestingspot(thecurrentrangesfrom1%Ibto4Ib),andyoushouldchooserightlyintherangeof1to8pulse.
Theindicationlightsofactiveandreactivearered.ThereactiveimportandexportcanbesetasⅠ+Ⅳ,Ⅱ+ⅢorⅠ+Ⅱ,Ⅲ+Ⅳuponrequestoftheuser.
1.5Transportationandstorage
1.5.1Themetershouldbefreeofvioletshockduringtransportationandstorage,andalsoshouldconformtoGB/T15464-1995Generaltechnicalconditionsforinstrumentandmeterpackaging
1.5.2Storagecondition:
thetemperatureshouldbe-40℃~70℃,relativehumidityshouldbelessthan85%andthereisnocorrosivegasexisting.Themetershouldbeplacedonthestandsinthewarehouse,withpilingupheightofindividualmeterlessthan5piecesandcartonlessthan5cartons
1.5.3Guaranteeperiod:
Ontheconditionofcorrectoperationaccordingtousermanualandthesealingiswithnodamage,themeterwillbegrantedfree-chargereparationorreplacementwithin3yearsifthemetercan’tmeetrequirementsstipulatedbytechnicalcondition
2.Function
2.1Meteringfunction
2.1.1Energymetering
a.Recordingandstoringactiveandreactive4quadrantcumulativetotalenergyinimportandexportforthismonth,lastmonthandthemonthbeforelastandetc.
b.Recording,displayingandstoringactiveandreactive4quadrantenergyateachtariff(sharp,peak,off-peak,shoulder)inimportandexportofthismonth,lastmonthandthemonthbeforelastandetc.
c.Recordingandstoringreactiveenergyineachquadrantandreactiveenergyabsolutevalueandthesumofpeak,off-peakandshoulderenergyofanytwoquadrantsforthismonth,lastmonthandthemonthbeforelast.
d.RecordingandstoringactivecumulativetotalenergyofphaseA,phaseBandphaseCinimportandexportforcurrentmonth,lastmonthandthemonthbeforelast.
e.Defaultdisplayingenergydataforlast3monthsfromthismonth,lastmonthtothemonthbeforelast(displaycontentcanbesetuponuser’srequest),storing12months’historicaldata
f.Energymeteringvalueis6integersand2decimals.TheunitiskWh,kVarhandkVAh.
2.1.2DemandMetering
a.RecordingandstoringimportandexportactiveMD,peak,off-peakandshoulderdemandaswellasthestateddemandoccurringdateandtimeofthismonth,lastmonthandthemonthbeforelast.
b.Recordingandstoringactivepeakandoff-peakMDanditsoccurringdateandtimeofthismonth,lastmonthandthemonthbeforelastmonth
c.RecordingandsavingreactiveMDanditsoccurringdateandtimeofthismonth,lastmonthandthemonthbeforelastmonth
d.Randomdisplaycurrentdemand,trulyreflectcurrentloadstatus
e.Whenthemeteroperatestothepre-setmeterreadingday(dateandtimecanbeset),itautomaticallyreadsMDandthenclears.MDclearingcanbeoperatedmanuallybypersonnelauthorized.
f.Defaultdisplaydemanddata(displaycontentcanbeset)oflast3months:
thismonth,lastmonthandthemonthbeforelast,andsavehistoricaldataof12months
g.Demandmeteringvalueis2digitsintegral,4decimals,andtheunitiskW,kvar,kVA
2.1.3Voltage,Current,PowerandFrequencyMetering
a.Real-timedisplayvoltage,currentofphaseA,BandC.
b.Real-timedisplayapparentpower,reactiveandactivepowerofphaseA,BandC.
c.Real-timedisplaynetworkfrequency
2.1.4PowerFactorMetering
Randomdisplaytotalpowerfactor,phaseA,BandCpowerfactorofcurrent15min.
2.2Voltageloss,currentlossandphaselossalarm,displayandrecordfunction
2.2.1WhenthecurrentI≥0.5%Ib(settable),andany(two)of3phaseisvoltageloseorlowerthan78%(settable)ratedvoltage,themeterisidentifiedasvoltageloss.Themeterwillsendsound-lightalarmanddisplaythephasewithfailure.Ifvoltagelossisover30sec(settable),themeterstartsinteriorvoltagelosslogprogramtorecordthephasesufferingloss,voltagelossstartandendtime,voltagelossaccumulativetimeandtimes,normalphaseenergyduringthetimeoffailurestartandend.Aftervoltagerecoversto85%(settable)ofratedvoltage,voltagelossalarmdisappearsandthemeterresumesnormaldisplayandmetering.(See4.5foralarmcode)
Whenthe3phasevoltagesuffervoltageloss,meterhasnodisplay,atthistime,ifthemeterhascurrentandI≥5%Ib,themeterwillbeidentifiedasnovoltage.Themeterwillrecordthistime’snovoltageaccumulativetimeandtimes,startandendtime.Andwhenthevoltageisnormal,itcandisplayaboverecordeddata.
2.2.2Currentloss
Whenthecurrentlowerlimit≥setvalue(settable),currentimbalancerate≥setvalue(settable),themeterisidentifiedascurrentloss.Ifcurrentlosssurpasses30sec(settable),themeterwillsendsound-lightalarm,displayphasewithfailure,startinteriorcurrentlosslogprogramtorecordthephaseo